1100 LD | 3100 | 5100 | 3350 | Boulder Counter | |
Specifications |
|||||
Size Range | 0.1 - 1.0 μm Optional ranges up to 5.0 µm |
0.3 - 25.0μm | 0.5 - 25.0μm | 0.3 - 25.0 µm | 5.0 - 100.0 um |
Channel Sizes | Standard 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0μm Optional 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 1.0, 5.0 µm |
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm Standard: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm 8-Channel: 0.3, 0.5, 0.7, 1.0, 3.0, 5.0,7.0,10.0 µm |
Standard: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 µm Standard: 0.5, 1.0, 3.0, 5.0, 10.0, 25.0 µm 8-Channel: 0.5, 0.7, 1.0, 2.0, 3.0, 5.0, 7.0,10.0 µm |
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm Optional: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm |
Standard: 5.0, 10.0, 25.0, 40.0, 50.0,100.0µm |
Flow Rate | 1.0 CFM (28.3 LPM) | 100 LPM | 28.3 LPM | ||
Counting Efficiency | 50% @ 0.1 µm; 100% for particles > 0.15 µm per ISO 21501-4 | 50% @ 0.3 µm; 100% @ sizes > 0.45 µm |
50% @ 0.5 um; 100% @ sizes > 0.75 µm |
50% @ 0.3 µm; 100% @ sizes > 0.45 µm | 50% @ 5.0 um |
Concentration Limits | 500,000 Particles / ft3 (17,150,000 / m3) @ 5% Coincidence Loss | 1,000,000 particles/ft3 (35,300,000/m3) @ 10% coincidence loss Suitable for ISO classes 1-8 cleanrooms |
400,000 particles/ft3 (14,120,000/m3) @ 10% coincidence loss Suitable for ISO classes 1-8 cleanrooms |
500,000 particles/ft3 @ 5% coincidence loss | |
Light Source | Long Life Laser Diode | Extreme Life Laser Diode Technology with >20-year Mean Time To Failure (MTTF) | |||
Dimensions | 14.65"Lx8.0"Wx10.4"H (37x20x26cm) | 9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm) | 14.65"Lx8.0"Wx10.4"H (37x20x26cm) | 9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm) | |
Weight | With battery: 23.55 lbs (10.68 kg) Without battery: 20.45 lbs (9.28 kg) |
With battery: 13.40 lbs (6.08 kg) Without battery: 12.00 lbs (5.44 kg) |
With battery: 26.15 lbs (11.86 kg) Without battery: 23.05 lbs (10.46 kg) |
With battery: 13.40 lbs (6.08 kg) Without battery: 12.00 lbs (5.44 kg) |
|
Zero Count Level | < 1 count / 5 minutes (per JIS B 9921) | ||||
Calibration | Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter | ||||
Count Modes | Automatic, Manual, Beep, Concentration, Cumulative/Differential, Real Time (with LMSNet or LMS Express RT software) | ||||
Data Storage | 3,000 Sample Records, Rotating Buffer (particle and environmental data with location and time) | ||||
Communication Modes | USB, Ethernet, RS-485 and Secure Data Transfer (LSDT) to and Storage on USB Flash Drive | ||||
Supporting Software | LMS Express and Express RT; LMS Pharma | ||||
Environmental Sensors | Optional (up to 4) temperature/relative humidity, air velocity, differential pressure or others | ||||
Touch Screen Display | 5.7-inch (14.47 cm) Color TFT Touch Screen Display | ||||
Reports | Pass/Fail Report Wizard for ISO-14644-1, EU GMP Annex 1 and FS-209 Report Formats | ||||
Alarms | Integrated alarm buzzer and external alarm connector. Alarms on counts, low battery, sensor failure | ||||
Sample Output | Internally Filtered to 100% @ 0.1μm, Optional HEPA Filtration for Cooling System Exhaust | Internally filtered to HEPA standards (>99.97% @ 0.3μm) | |||
Vacuum Source | Internal closed-loop controlled pump. Allows the use of 3/8" ID sample tubing of up to 12 meters (40 feet) @ 50 & 100LPM | ||||
Enclosure | Stainless Steel | ||||
Power | 100-240 VAC, 50-60 Hz, External AC/DC Adapter | ||||
Battery | Optional: Li-Ion, Removable & Rechargeable (typical recharge time 7.5 hours) | ||||
Languages | English (U.S.), French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean | ||||
Environmental Conditions |
|||||
Operating | 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing | ||||
Storage | 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing | ||||
Accessories |
|||||
Included | Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Power Cable and Spare Fuse; LMS XChange and Express Software; Printer; Printer Paper | ||||
Optional | Printed Operating Manual; Network Adapter; Wireless Network Adapter; Internal Removeable, Rechargeable Battery; Carrying Case with Wheels; Cleanroom Cart; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; Filter Scanning Probe; High Pressure Difuser; 32-Port Universal Manifold; LMS Express RT Software; LMSNet Software; Validation Documentation | Wireless Network Adapter; Removable Rechargeable Li-Ion Battery; Carrying Case with Wheels; Temperature/Relative Humidity Probe; Differential Pressure Probe; Intelligent Filter Scanning Probe; Air Velocity Probe; Spare Removable Rechargeable Li-Ion Battery; External Battery Charger; LMS Express, RT Software; LMS Pharma Edition Software; LMS Standard Software; Validation Documentation; Printed Operating Manual; Cleanroom Cart |
Solair 1100/3100/5100 Tutorial Series | Configure Sample Settings
Solair 1100/3100/5100 Tutorial Series | Configure Alarm Settings
Solair 1100/3100/5100 Tutorial Series | Configure Security Settings
Reports Function
Locations Function
The SOLAIR 1100LD (Laser Diode) leads the industry as the first of its kind portable particle counter with 0.1-micron sensitivity and 1.0 CFM (28.3 LPM) flow rate using laser diode technology. Conventional HeNe (Helium Neon) particle counters require frequent cleaning, periodic Laser tube replacement and suffer low concentration limits. The SOLAIR 1100LD is designed to operate in ISO Class 1 to Class 7 cleanrooms without concerns about degrading its environment or exceeding concentration limits.
Number 1 in Versatility
The SOLAIR line of particle counters is the most versatile on the market today.
Whether you’re certifying your cleanroom to ISO standards, performing routine testing, counting particles in your compressed gas lines, scanning your HEPA filters for leaks, challenge testing your HEPA filters, or measuring environmental parameters such as TRH, DP, or AV, the SOLAIR portable particle counter is the right instrument to get the job done.
Cleanroom Certification
Whether you certify current ISO 14644 standards, EU GMP Annex 1, or Federal Standards 209E, the SOLAIR has you covered with its certification report wizard.
Routine Testing
Routine testing is a breeze with the intuitive interface, zoom screen, and onboard sample recipes.
Troubleshooting
With beep and concentration sample modes, the SOLAIR will simplify the identification of particle sources within your cleanroom.
Compressed Gas Testing
The SOLAIR can test for particle counts in compressed air by utilizing a high-pressure diffuser or high-pressure controller. Particles can be tested down to 0.1 microns.
HEPA Filter Scanning
The scanning of HEPA filters is a critical step to ensure the integrity of the HEPA filter and the overall cleanliness of the cleanroom. The ScanAir attachment from Lighthouse makes filter scanning a more effortless operation.
HEPA Filter Challenge Testing
Monitoring of other environmental parameters such as Temp, Humidity, Differential Pressure or Air Velocity
Other Environmental Testing
The SOLAIR can record data from up to 4 environmental sensors that plug into and are powered by the SOLAIR. Data is recorded into the 3,000 record data buffer and can be downloaded along with the particle count data to your computer via USB key.
Better Technology Leads to Greater Flexibility
The SOLAIR 1100 uses a laser diode as its laser source. This laser technology provides the SOLAIR 1100 with a 10x increase in concentration limits relative to the competition. The higher concentration limit allows the instrument to be used in ISO Class 1 to Class 7 environments.
Save Time
Spend less time configuring your particle counter and more time counting particles. The SOLAIR features one of the industry’s best and easy-to-use touch screen interfaces (UI). Sample configuration is a breeze with ICON-driven menus, straightforward layout, and enhanced responsiveness.
The intuitive interface reduces learning uptime for new users and simplifies the writing of SOP’s for operation.
The SOLAIR has an integrated report wizard for cleanroom certification that makes cleanroom certification a breeze. Once the basic sample parameters are input, the wizard will prompt the end-user to move to the following sample location, which helps to ensure no sample locations are missed.
The configurable recipe database can store up to 50 recipes for sampling and reports. These recipes reduce operator configuration errors as well as setup time.
Ease of Use
Lighter is better, and the SOLAIR 1100 is the lightest 0.1-micron particle counter on the market, at only 23 lbs. without battery and 26 with a battery.
The SOLAIR can be powered via an onboard battery or plugged directly into a power outlet for extended use. With the fastest battery recharge time of any portable on the market, the SOLAIR offers unrivaled uptime.
With beep and concentration sample modes,
the SOLAIR will simplify the identification of particle sources within your cleanroom.
Data from the SOLAIR can be downloaded directly to a standalone software package or a real-time monitoring system from Lighthouse.
Flexible communications with these instruments are made possible by the integrated Ethernet, RS-485, USB, and USB flash drive interfaces.
Reliability
The line of SOLAIR portable particle counters incorporate laser diode technology.
The vacuum system within the SOLAIR was designed with closed-loop feedback to ensure accurate sample volumes. The vacuum pump was designed and built by Lighthouse to ensure long-term reliability.
The SOLAIR 1100LD is designed, manufactured, and calibrated to ISO 21501-4 to ensure particle-counting accuracy.
The SOLAIR can store up to 3,000 records of particle count data from up to 8 particle channels, and the configurable recipe database can store up to 50 recipes for sampling and reports. All data can be quickly and reliably transferred to a computer, a Facility Monitoring System or a Building Automation System. Secure data transfer (per 21 CFR Part 11) to a USB flash drive can be accomplished using the Lighthouse secure data transfer feature and LMS Express software.
Key Features:
SERVICE & CALIBRATION